HomeAbout UsNews & EventsSupport Center Media Relations grace.com > DiscoverySciences.com
PRINT PAGE E-MAIL PAGE

Vancomycin Hydrochloride

Vancomycin Hydrochloride
Column:   Vydac® 218TP C18, 5µm, 3.2 x 250mm
(Part No. 218TP53)
Mobile  
Phase:  
TEA buffer = 4mL of Triethylamine in 2000mL Water, adjusted to pH 3.2 with Phosphoric Acid
Gradient:   Hold 92:7:1 TEA buffer:Acetonitrile:THF for 12 minutes, then ramp linearly to 70:29:1 over 8 min
Detector:   UV at 280nm
Flow rate:   1.0mL/min
Column  
Temp:  
Comments:   USP requires R not less than 3.0. The high resolution of the 218TP column splits resolution peak 1 into two peaks. The second of these, at 4.7 minutes and closest to the vancomycin peak, was used to calculate the R value.
1.  Res. Cpd. 1
2.  Vancomycin
3.  Res. Cpd. 2